点击这里给我发消息点击这里给我发消息
联系我们
服务热线
0572-8832711
公司地址:湖州市德清县雷甸镇运河路58号4号楼3层
ET2710 Digital Resistance RTD TC Temperature Calibrator
浏览: 发布日期:2018-07-06


ET2710 Digital Resistance RTD TC Temperature Calibrator
 
Features and Benefits
· It allows you to measure DC V, DC mV, resistance ,TC and RTD.
· Outputs resistance, simulation of TC and RTD.
· Automatic and manual cold end temperature compensation for TC.
· Manual stepping, automatic stepping, 0~100% phase step and ramp output.
· Super-strong protection:The waterproof grade is IP67. Automatic protection against the   signal terminals connect to the 220V.
Summary of Source and Measure Functions
Function Measure Source
DC V 0~30V Not available
DC mV 0~120mV 0mV~100mV
Resistance 0~3200Ω 0~3200Ω
RTD Pt100, Pt1000, Cu50, Cu100 Pt100, Pt1000, Cu50, Cu100
TC E,J,K,T,B,R,S,N E,J,K,T,B,R,S,N
Others Stepping output, ramp, phase step output, user-defined range
 
  Specifications
 
Specifications are based on a one year calibration cycle and apply from +18°C to +28 °C unless stated otherwise. All specifications assume a 10 minute warmup period.
DC Voltage Measurement
Range Maximum measurement range Resolution  
Accuracy
 (% of reading+counts)
ET2710 (0.05%) ET2710 (0.02%)
30V 0V~31V 0.001V 0.05+2 0.02+2
100mV -15mV~125mV 0.01mV 0.05+2 0.02+2
-10°C~18°C,+28°C~55°C temperature coefficient,
±0.005%FS/°C.Input resistance: >1MΩ.
  Resistance Measurement
Range Maximum measurement range Resolution Accuracy (Ω)
ET2710 (0.05%) ET2710 (0.02%)
2-wire, 3-wire 4-wire 2-wire,
3-wire
4-wire
400Ω 0~440Ω 0.01Ω 0.25 0.15 0.15 0.10
3200Ω 420Ω~3600Ω 0.1Ω 1.5 1.0 1.0 0.5
-10°C~18°C,+28°C~55°C temperature coefficient,±0.005%FS/°C.
Exciting current during measurement:
400Ω:1.0mA±10%;
3200Ω:0.2mA±10%;
Two-wire: Conductor resistance is excluded from errors.
Three-wire: Matching test line should be used.
The total resistance of conductor should not be larger than 25Ω.
   Temperature, TC
Graduation Range Resolution Accuracy (Ω)
ET2710 (0.05%) ET2710 (0.02%)
J -200°C~0°C
0°C~1200°C
0.1°C 1.5°C
1.0°C
1.0°C
0.7°C
K -200°C~0°C
0°C~1370°C
0.1°C 1.8°C
1.2°C
1.2°C
0.8°C
T -200°C~0°C
0°C~400°C
0.1°C 1.8°C
1.2°C
1.2°C
0.8°C
E -200°C~0°C
0°C~950°C
0.1°C 1.5°C
1.0°C
0.9°C
0.7°C
R -20°C~0°C
0°C~500°C
500°C~1750°C
1°C 4°C
2.5°C
2°C
2.5°C
1.8°C
1.4°C
S -20°C~0°C
0°C~500°C
500°C~1750°C
1°C 4°C
2.5°C
2°C
2.5°C
1.8°C
1.5°C
B 600°C~800°C
800°C~1000°C
1000°C~1800°C
1°C 3.5°C
2.5°C
2°C
2.2°C
1.8°C
1.4°C
N -200°C~0°C
0°C~1300°C
0.1°C 2.0°C
1.2°C
1.5°C
0.9°C
Errors of cold-junction compensation are not included in the table.
Accuracy of cold-junction compensation: 1.5°C
  Temperature, RTD
The calibrator accepts Pt100, Pt1000, Cu50 and Cu100, etc. The calibrator accepts RTD measurements inputs in two-, three-, or four-wire connections, with the three-wire connection the most common. A four-wire configuration provides the highest measurement precision, and two-wire provides the lowest measurement precision.
 
 
 
 
 
The other parameters
Temperature: - 10 ℃ ~ 55 ℃
Storage temperature: - 20 ℃ ~ 60 ℃
Weight: 350 grams
Dimensions: 185mm*93mm*47mm
Power supply: 6Pcs AAA battery ni-mh or AC ADAPTER
Communication channel: RS232
 
18P02 communication wires
Power adapter(18A52)
Probes(18A53)
 
 

Copyright © 2017-2020 杭州中创电子有限公司     备案号:浙ICP备18021705号-4

技术支持:杭州中创