ET2712 RTD Calibrator
For the temperature calibration professional that wants a highly accurate, easy-to-use, single function RTD temperature. It is a kind of calibrating RTD temperature transmitter instrument. It can measure and simulate various type RTD. It can also simulate and test resistance. calibrator the ET2712 is the ideal test tool.
Features
Accuracy:0.02% or 0.05%
Measure output temperature of thermal resistance
May simulate output of thermal resistance
Automatic identification of three-wire or four-wire connection mode
Perfect performance on waterproof protection: IP67 grade
In case of connecting 220V with automatic over voltage protection function.
With manual stepping, automatic stepping and 0 ~ 100% output of phase step and ramp function
LED with white backlight
Summary of Source and Measure Functions
Function | Measure | Source |
DC V | 0~30V | Not available |
Resistance | 0~3200Ω | 0~3200Ω |
RTD | Pt100, Pt1000, Cu50, Cu100 | Pt100, Pt1000, Cu50, Cu100 |
Others | Stepping output, ramp, phase step output, user-defined range |
Specifications are based on a one year calibration cycle and apply from +18°Cto +28 °C unless stated otherwise. All specifications assume a 10 minute warmup period.
1,DC Voltage Measurement
Range | Maximum measurement range | Resolution | Accuracy (% of reading + Count) | |
ET2712( 0.05%) | ET2712( 0.02%) | |||
30V | 0V~31V | 0.001V | 0.05+2 | 0.02+2 |
-10°C~18°C,+28°C~55°C temperature coefficient,±0.005%FS/°C. Input resistance: >1MΩ. |
Range | Maximum measurement range | Resolution | Accuracy (Ω) | |||
ET2712( 0.05%) | ET2712( 0.02%) | |||||
2-wire, 3-wire | 4-wire |
2-wire, 3-wire |
4-wire | |||
400Ω | 0~440Ω | 0.01Ω | 0.25 | 0.15 | 0.15 | 0.10 |
3200Ω | 420Ω~3600Ω | 0.1Ω | 1.5 | 1.0 | 1.0 | 0.5 |
-10°C~18°C,+28°C~55°C temperature coefficient,±0.005%FS/°C. Exciting current during measurement: 400Ω:1.0mA±10%; 3200Ω:0.2mA±10%; Two-wire: Conductor resistance is excluded from errors. Three-wire: Matching test line should be used. The total resistance of conductor should not be larger than 25Ω. |
Range | Maximum output range | Resolution | External exciting current | Accuracy (Ω) | |
ET2712( 0.05%) | ET2712( 0.02%) | ||||
400Ω | 0~440Ω | 0.01Ω | 0.4mA~4.0mA | 0.25 | 0.15 |
3200Ω | 400~3600Ω | 0.1Ω | 0.1mA~0.5mA | 1.0 | 0.50 |
-10°C~18°C,+28°C~55°C temperature coefficient,±0.005%FS/°C. |
Graduation | Range | Resolution | Accuracy (°C) | |||||
ET2712( 0.05%) | ET2712( 0.02%) | |||||||
2-wire 3-wire |
4-wire | Output |
2-wire 3-wire |
4-wire | Output | |||
Pt100 | -200°C~850°C | 0.1°C | 0.7 | 0.4 | 0.7 | 0.4 | 0.3 | 0.3 |
Pt1000 | -200°C~650°C | 0.4 | 0.3 | 0.3 | 0.3 | 0.15 | 0.15 | |
Cu50 | -50°C~150°C | 1.2 | 0.8 | 0.8 | 0.8 | 0.5 | 0.5 | |
Cu100 | -50°C~150°C | 0.7 | 0.4 | 0.4 | 0.4 | 0.25 | 0.25 | |
As for exciting current during measurement, please refer to resistance measurement function. As for allowable external exciting current during output, please refer to resistance output function. 2-wire: Does not include lead resistance. 3-wire: Assumes matched leads with a total resistance not exceeding 25Ω. |
Using environment temperature:-18℃~55℃
Store environment temperature:(-20 – 60)°C
Relative humidity :0~ 90RH No condensation
Weight:0.35Kg
Size:185mm×93mm×47mmm
Power supply: 6 R03-model rechargeable batteries or power adapter
communication mode:RS232
1,Standard Accessories
A set of calibrator also includes the following items:
hard spot test leads (one set)
alligator clips (one set)
One 12V/1A power adaptor
Users Manual
2,Optional Accessories
6 R03-model rechargeable batteries
1 Metal Box
18P02 communication wires
Power adapter(18A52)
Probes(18A53)