ET2715 Current and Voltage Calibrator
Summary:
ET2715 Current and Voltage Calibrator is a handheld calibration instrument with high precision, high resolution and high reliability that integrates output and measurement of V, mV, mA and Hz signals. The meter adopts large liquid crystal display with clear reading, and it also has adjustable back light source and automatic shutdown function. The meter can be supplied by dc or ac power supply, which is more convenient to use.
Features
Accuracy:0.02% or 0.05%
Perfect performance on waterproof protection: IP67 grade
In case of connecting 220V with automatic over voltage protection function.
With manual stepping, automatic stepping and 0 ~ 100% output of phase step and ramp function
LED with white backlight
Source of voltage, current and frequency.
Measure of voltage, current, loop current and frequency.
Manual stepping and automatic stepping and ramping.
Summary of Source and Measure Functions
Function | Measure | Source |
DC V | 0~30 V | 0~10 V |
DC mV | 0~100 mV | 0~100 mV |
DC mA | 0~24 mA | 0~24 mA |
Frequency | 1.000Hz~99.999kHz | 0.00Hz~20.000kHz |
Others | 24V power supply, Step, Ramp. |
Specifications are based on a one year calibration cycle and apply from +18°Cto +28 °C unless stated otherwise. All specifications assume a 10 minute warmup period.
1,DC Voltage Measurement
Range | Maximum measurement range | Resolution |
Accuracy (% of reading + Counts) |
|
ET2715(0.05%) | ET2715(0.02%) | |||
30V | 0V~31V | 0.001V | 0.05+2 | 0.02+2 |
100mV | -15mV~125mV | 0.01mV | 0.05+2 | 0.02+2 |
-10°C~18°C,+28°C~55°C temperature coefficient,±0.005%FS/°C. Input resistance: >1MΩ. |
Range | Maximum measurement range | Resolution |
Accuracy (% of reading + Counts) |
|
ET2715(0.05%) | ET2715(0.02%) | |||
20mA | 0~24mA | 0.001mA | 0.05+2 | 0.02+2 |
-10°C~18°C,+28°C~55°C temperature coefficient,±0.005%FS/°C. Input resistance:<100Ω. |
Range | Maximum measurement range | Resolution |
Accuracy (% of reading + Counts) |
|
ET2715(0.05%) | ET2715(0.02%) | |||
100Hz | 1~99.999Hz | 0.001Hz | 0.02+1 | 0.01+1 |
1000Hz | 100~999.99Hz | 0.01Hz | ||
10kHz | 1k~9.9999kHz | 0.1Hz | ||
100kHz | 10k~99.999kHz | 1Hz | ||
Sensitivity: 10Hz~10kHz,Vp-p≥1V; the rest:Vp-p≥2V. Wave form: 5 counting points should be added to errors of other wave forms. Commercial frequency can be measured directly. |
Using environment temperature:-18℃~55℃
Store environment temperature:(-20 – 60)°C
Relative humidity :0~ 90RH No condensation
Weight:0.35Kg
Size:185mm×93mm×47mmm
Power supply: 6 R03-model rechargeable batteries or power adapter
communication mode:RS232
1,Standard Accessories
A set of calibrator also includes the following items:
hard spot test leads (one set)
alligator clips (one set)
One 12V/1A power adaptor
Users Manual
2,Optional Accessories
6 R03-model rechargeable batteries
1 Metal Box
18P02 communication wires
Power adapter(18A52)
Probes(18A53)